MINI-SED is a scintillator type detector that captures secondary electrons and outputs it as an electrical signal. It is mainly used as a detector for scanning electron microscopes.


We will respond to your request of special specifications.


MINI-SED va=2000V va=200V va=30V
MINI-SED Secondary electron images
SampleFCu 400 meshiL•SF32ƒΚ‚j

Features


Compact size

It can be mounted in a small space. (Interface: ICF 70)

Ultrahigh vacuum system

Baking up to 150 Ž is possible. It can be used at a vacuum degree less than 1~10-2.

Maximum response speed: 5MHz

Current detection range is from pA to ƒΚA. (Maximum amplification: 2~106 times, Output: 1 Vp-p)


Customized SED


with shutter, ICF 114 Short type (length: 50 mm) Thin type (diameter: ƒΣ10 mm)


Specification


Quantum conversion method

Scintillator

Configuration

Scintillator

P47

Photomultiplier tube

R9880U / Hamamatsu Photonics K.K.

Light Guide

Quartz glass

Pre-amplifier

V-C converter, Impedance: 20 kƒΆ

Range of detection

Current amount

1 pA`1 ƒΚA

Response speed

5 MHz

Maximum gain of PMT

2~106

Output voltage

0`1 Vp-p

Input

to Preamplifier

DC}8 V / 30mA

to PMT

0`-1 kV / 300 ƒΚA

to Scintillator

0`+8 kV / 10 ƒΚA

Condition

Degree of vacuum

1~10-2 Pa

Baking temperature

Less than 150 Ž (Except the preamp)

Interface

ICF 70

Dimensional drawing (example)



*The power supply is not included.


Factory-recommended power supply


Power supply for secondary electron detector is also on sale.


Functions

Brightness, Contrast, Filter

Input

fot Pre-amplifier (voltage)
for PMT (voltage)
for Scintillator (voltage)

Output